Article,

Amperometric immunosensors based on rigid conducting immunocomposites.

, , , and .
Anal Chem, 69 (11): 2080--2085 (June 1997)
DOI: 10.1021/ac961222b

Abstract

Novel polishable immunosensors based on rigid biocomposite materials have been constructed. These biocomposites contain graphite powder, rabbit IgG, and methacrylate or epoxy resins. This material acts as a reservoir for the biological molecules and as a transducer at the same time. In order to study the potential analytical properties of this new type of material, a competitive binding assay was developed to determine the RIgG present in a sample with the aid of goat anti-rabbit IgG labeled with alkaline phosphatase. Using phenyl phosphate as a substrate, the phenol produced by the enzymatic reaction was amperometrically detected at 800 mV (vs Ag/AgC1). The surface of the immunosensor can be regenerated by simply polishing, obtaining fresh immunocomposite ready to be used in a new competitive assay.

Tags

Users

  • @biblio24

Comments and Reviews