A simple handheld push-button device for in situ calibration of pneumotachographs

, , , and .
J Appl Physiol, 77 (4): 2042--2047 (October 1994)

Abstract

A simple and compact flow calibrator has been devised for generating precise predetermined constant flow rates for checking the calibration of laboratory and clinical flow transducers used in respiratory measurements. The standard version delivers preset flows of 0.5 and 1 l/s, whereas a tuned-up version can produce preset flows of 2.5 and 5 l/s, with an accuracy of +/- 2%. The pressure generated is sufficient to cope with most commonly used respiratory flowmeters. The flow calibrator is built from inexpensive components that are readily obtainable: a fan, a turbine flowmeter, and a feedback circuit in a compact housing. The device is easy to connect to other equipment and to operate. Three flow calibrators have been built and are in regular use in a lung function laboratory and on intensive care wards.